Analysis reveals significant concordance between mass spectrometry and immunofixation electrophoresis for detecting therapeutic antibodies in multiple myeloma patients.
Background/Objectives: Therapeutic monoclonal antibodies, including bispecifics (t-mAbs), can interfere with serum protein electrophoresis (SPEP) and immunofixation electrophoresis (IFE), mimicking residual M-protein. We evaluated a mass spectrometry (MS; EXENT®)-based workflow supported by an m/z reference library to discriminate drug from disease and assess concordance with IFE. Methods: Fifty-eight serum samples from 29 multiple myeloma patients were analyzed at baseline and after 3 months. Targeted enrichment of t-mAbs followed by MS enabled detection of peaks annotated through matching to a theoretical m/z panel and correlation with SPEP/IFE results. Results: Comparison of IFE versus MS showed 11/29 (38%) double positives, 15/29 (52%) double negatives, and 3/29 (10%) IFE−/MS+; no IFE+/MS− cases were observed. Using MS as a reference, IFE exhibited 78.6% sensitivity and 100% specificity. The m/z library enabled attribution of interference to linvoseltamab (n = 9), daratumumab (n = 6), and teclistamab (n = 3); in 16 patients treated with other bispecifics, no drug-related peaks were detected after 3 months. Longitudinal analysis discriminated therapeutic from endogenous immunoglobulins, identified baseline M-protein, and prevented false residual signals. Conclusions: MS (EXENT®)-based characterization of t-mAbs improves response monitoring accuracy in multiple myeloma and supports integration of MS into routine laboratory practice.
No takes yet. Share an insight, caveat, or question.
Iglesias et al. (2025) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: