Case report reveals pseudo-right bundle branch block during ventricular pacing in a 67-year-old male, highlighting physiological septal conduction rather than lead malposition.
Background Right ventricular (RV) pacing typically produces left bundle branch block (LBBB) morphology. Right bundle branch block (RBBB) patterns during RV lead placement raise concerns for malpositioning. Case Presentation A 67‐year‐old male with symptomatic atrioventricular dysfunction and bi‐fascicular block underwent dual‐chamber pacemaker implantation. During RV lead testing, unexpected RBBB morphology was noted despite correct positioning. Results Fluoroscopic verification excluded malpositioning. The Pseudo‐RBBB pattern most likely resulted from physiological conduction through existing right bundle disease, where impulses travelled preferentially via septal pathways to activate the left ventricle. Conclusion RBBB morphology during RV pacing requires systematic evaluation to exclude malpositioning but may represent physiological conduction in patients with or without existing conduction disease.