Answers are grounded in this paper and cite their sources.
Analysis reveals glucometer inaccuracies in clinical decision-making at tertiary care hospital, suggesting gaps in performance.
Our study demonstrates that a substantial gap exists in the analytical performance of the glucometer compared to the standard laboratory analyser when evaluated against ISO 15197:2013 criteria. However, error grid analysis suggests that the majority of discrepancies observed had minimal impact on clinical decision-making in our study setting.
No takes yet. Share an insight, caveat, or question.
Khan et al. (2025) studied this question.