Technique improves phase stabilisation in optical coherence tomography, suggesting reduced phase variations.
This work presents a technique designed to correct laser-induced phase variations in Swept-Source optical coherence tomography (SS-OCT) instruments. Conventionally, data resampling is achieved by using an external k-clock module, and the signal generated by taping some of the light and diverting it through a fibre Bragg grating is used to trigger the acquisition and stabilise the phase. Instead, in this work, the nonlinearities in the imaging interferometer output are computed from data generated by a reference interferometer block in synchronism with data collected by the imaging interferometer. Phase-stabilised digital channelled spectra are obtained by monitoring the number of cycles in the output signal of the reference interferometer. Our preliminary results show a significant improvement in the stabilisation of the phase, from an initial 3.6 radians to less than 0.1 radians when using our approach.
No takes yet. Share an insight, caveat, or question.
Jiménez et al. (2025) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: