Key result
Perceval PLUS shows improved hemodynamics and lower PPI rates versus Perceval S.
Why the study?
Does the Perceval PLUS sutureless aortic valve improve hemodynamics and clinical outcomes compared to the Perceval S design in patients undergoing aortic valve replacement?
Cohort (n=1,136)
Does the Perceval PLUS sutureless aortic valve improve hemodynamics and clinical outcomes compared to the Perceval S design in patients undergoing aortic valve replacement?
The newer Perceval PLUS sutureless aortic valve provides improved hemodynamics and low pacemaker implantation rates with similar early mortality compared to the older Perceval S design.
Answers are grounded in this paper and cite their sources.
May guide valve choice in surgical AVR; extends observational data on Perceval platform performance.
Introduction In 2020, the sutureless Perceval PLUS aortic valve was introduced, featuring a new anticalcification treatment (FREE) and a shorter inflow ring in the XL size. We aimed to evaluate its early outcomes compared to earlier Perceval S design.Methods We retrospectively analyzed 1136 patients who received the Perceval bioprosthesis between 2007 and 2022, including 222 implanted with the Perceval PLUS. Patients implanted with the Perceval S were divided into two groups (S-OLD and S-NEW) depending on the sizing strategy. Survival, hemodynamic performance, and pacemaker implantation rates were assessed.Results The cohort had a mean age of 77 years and a EuroSCORE II of 6.3%. The observed 30-day mortality was similar across groups (3.7%, p = .326). Perceval PLUS demonstrated significantly improved hemodynamics, with lower peak (20.44 ± 8.95 mmHg) and mean gradients (11.44 ± 5.12 mmHg) compared to the S design (p < .001). A permanent pacemaker implantation (PPI) rate of 5.3% was achieved in minimally invasive aortic valve replacement in the PLUS group.Conclusion Early results from the Perceval PLUS show a secure profile with maintained benefits from the Perceval platform. The Perceval PLUS shows improved hemodynamics in comparison with the Perceval S and achieves low PPI rate.
No takes yet. Share an insight, caveat, or question.
Szecel et al. (2026) conducted a cohort in Aortic valve replacement (n=1,136). Perceval PLUS aortic valve vs. Perceval S aortic valve was evaluated on 30-day mortality (p=.326). The Perceval PLUS aortic valve demonstrated similar 30-day mortality compared to the Perceval S design but improved hemodynamics with lower peak and mean gradients.
Synapse has enriched 10 related papers. Consider them for comparative context: