Aortic hemodynamic indices, such as WDI and LDI, showed significantly higher scores in patients with Structural Valve Deterioration, indicating potential predictive value for SVD risk.
Do personalized computational hemodynamic indices correlate with the development of structural valve deterioration in patients after transcatheter aortic valve implantation?
Personalized computational hemodynamic indices can identify TAVI patients at high risk for structural valve deterioration, potentially guiding patient-specific follow-up strategies.
Absolute Event Rate: 0% vs 0%
Introduced as an alternative to open-heart surgery for elderly patients, Transcatheter Aortic Valve Implantation (TAVI) has recently been extended to younger patients due to comparable performance with the gold-standard. However, the long-term durability of the bio-prosthetic TAVI valves is limited by Structural Valve Deterioration (SVD), an inevitable degenerative process whose pathogenesis is still unclear. In this study, we aim to computationally investigate a possible relationship between aortic hemodynamics and SVD development. To this aim, we collect data from twelve patients with and without SVD at long-term follow-up exams. Starting from pre-operative clinical images, we build early post-operative virtual geometries and perform Computational Fluid Dynamics simulations by prescribing a personalized flow rate based on Echo Doppler data. In order to identify a premature onset of SVD, we propose three computational hemodynamic indices: Wall Damage Index (WDI), Leaflet Delamination Index (LDI), and Leaflet Permeability Index (LPI). Additionally, to each index we associate a score and, using the Wilcoxon rank-sum test, we find that each score individually shows a statistically greater median value in the SVD sub-population (WDI: p=0.008, LDI: p=0.001, LPI: p=0.020). Finally, we define a synthetic scoring system that clearly separates between SVD and non-SVD patients. Our results suggest that aortic hemodynamics may drive a premature onset of SVD, and the synthetic score could potentially assist clinicians in a patient-specific planning of follow-up exams to closely monitor those patients at high SVD risk.
Crugnola et al. (Mon,) reported a other. Aortic hemodynamic indices, such as WDI and LDI, showed significantly higher scores in patients with Structural Valve Deterioration, indicating potential predictive value for SVD risk.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: