PulseExploreJournal ClubDebatesTrendingResearchersJournals
Instagram
HomeExploreJournal ClubTrending
Synapse
⌘+K
Synapse
January 10, 2026Journal of Hypertension0 citations

Performance of an automated Korotkoff sound BP monitor in cardiovascular patients

View Full Paper
PYPo-Kai YangKLKun-I LinPTPei-Yuan Tsai

Key Result

The Ksens-BP monitor showed excellent agreement with manual measurements (SBP CCC = 0.952) and better accuracy than Oscillo-BP in cardiovascular patients.

Key Points

  • To evaluate the accuracy of the Ksens-BP monitor compared to traditional blood pressure measurement methods in cardiovascular patients.
  • Single-center prospective observational study
  • Enrolled adults with cardiovascular disease
  • Conducted three paired BP measurements with Ksens-BP, Oscillo-BP, and Auscl-BP
  • Assessed agreement using concordance correlation coefficient (CCC)
  • Analyzed effects of comorbidities on measurement differences with hierarchical linear models
  • Ksens-BP showed excellent agreement with Auscl-BP (SBP CCC = 0.952; DBP CCC = 0.945)
  • Ksens-BP had smaller mean absolute differences compared to Oscillo-BP
  • Oscillo-BP accuracy diminished in patients with peripheral artery disease and atrial fibrillation, while Ksens-BP remained unaffected
  • 178 patients provided 686 valid paired measurements

Structured PICO

Does an automated Korotkoff sound-based monitor improve blood pressure measurement accuracy compared to oscillometric monitors in patients with cardiovascular disease?

P
Population
178 adults with cardiovascular disease (eligible conditions included hypertension, diabetes, atrial fibrillation, myocardial infarction, coronary artery disease, peripheral artery disease, stroke, or heart failure), mean age 67.4 years, 80% men.
I
Intervention
Automated Korotkoff sound-based monitor (Ksens-BP) integrating a semiconductor strain-gauge sensor and artificial intelligence waveform classification
C
Comparator
Oscillometric monitor (Oscillo-BP) and manual auscultatory method (Auscl-BP)
O
Outcome
Agreement with manual auscultatory blood pressure (Auscl-BP) assessed by concordance correlation coefficient (CCC)surrogate

A novel automated Korotkoff sound-based monitor provided more accurate blood pressure measurements than standard oscillometric devices in patients with cardiovascular disease, remaining robust even in the presence of atrial fibrillation and peripheral artery disease.

Abstract

Objectives: Accurate blood pressure measurement is essential for cardiovascular risk management, but conventional oscillometric devices are unreliable for atrial fibrillation and arterial stiffness. We evaluated the accuracy of a novel automated Korotkoff sound-based monitor (Ksens-BP), integrating a semiconductor strain-gauge sensor and artificial intelligence waveform classification, compared to oscillometric (Oscillo-BP) and manual auscultatory (Auscl-BP) methods. Methods: This single-center prospective observational study enrolled adults with cardiovascular disease at National Cheng Kung University Hospital, Douliu Branch (October 2023–January 2024). Eligible conditions included hypertension, diabetes, atrial fibrillation, myocardial infarction, coronary artery disease, peripheral artery disease, stroke, or heart failure. Participants underwent three paired blood pressure (BP) measurements with Ksens-BP, Oscillo-BP, and Auscl-BP using a unified cuff system. Agreement was assessed with Auscl-BP assessed by concordance correlation coefficient (CCC). Hierarchical linear models (HLMs) examined the effects of comorbidities on measurement differences. Results: A total of 178 patients (mean age 67.4 years; 80% men) contributed 686 valid paired measurements. Ksens-BP demonstrated excellent agreement with Auscl-BP (SBP CCC = 0.952; DBP CCC = 0.945) compared with Oscillo-BP (SBP CCC = 0.903; DBP CCC = 0.851). Mean absolute differences were smaller with Ksens-BP than Oscillo-BP (SBP: 2 vs. 4.4 mmHg; DBP: 2.3 vs. 5.4 mmHg). Oscillo-BP accuracy was negatively affected by PAD and atrial fibrillation, whereas Ksens-BP performance was unaffected by comorbidities. Conclusion: The Ksens-BP system demonstrated superior accuracy and robustness, providing reliable BP measurement across complex cardiovascular populations.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Yang et al. (2025) studied this question. The Ksens-BP monitor showed excellent agreement with manual measurements (SBP CCC = 0.952) and better accuracy than Oscillo-BP in cardiovascular patients.

synapsesocial.com/papers/6963221091e05aa366cb8844https://doi.org/10.1097/hjh.0000000000004232