Demonstrates improved beam monitoring in a low-intensity proton beamline, suggesting enhanced diagnostic capabilities.
We have developed a highly sensitive beam profile monitor, the High-Gain Residual Gas Ionization Profile Monitor (HGRGIPM). The HGRGIPM detects electrons ionized by the proton beam in residual gas, which are guided by electric fields to a phosphor screen. The fluorescence is collected by an optical system. RGIPMs have proven to be powerful diagnostic tools for high-intensity beams because they have non-destructive nature. They have been used as the main profile monitors in the primary beamline, A-line (30 GeV, 83 kW, vacuum ~0.3 Pa). However, the newly constructed B-line (11 W) presents a significant challenge, as its beam intensity is four orders lower than that of the A-line. To overcome this issue, several improvements were made in the design of the HGRGIPM. These include the use of a thin phosphor screen to enhance the signal-to-noise ratio, optimizing electrode geometry for higher acceleration voltages and improved sensitivity, and incorporating a high-efficiency optical system. These innovations allow profile monitoring even in the low-intensity B line. In this presentation, we report on the HGRGIPM's design and performance, beam response measurements, and future prospects.