Feasibility study demonstrates validity of a novel knee brace for monitoring function in daily activities, suggesting potential for rehabilitation.
Tracking knee joint movement during activities of daily living can have the potential to transform the rehabilitation and functional assessment of patients. The present study evaluated the validity of a low-cost, instrumented knee brace to determine whether it was appropriate for the monitoring and quantification of human knee function during five activity-of-daily-living (ADL) tasks including walking, inclined walking, stepping, sitting, and object manipulation. A sensor platform was designed to acquire sagittal plane knee data from 13 healthy participants across five different tasks and compared to gold-standard motion analysis. The brace showed good-to-excellent validity (RMSE: 4.97–8.65°), with differences in knee joint angles and angular velocities noted during various ADLs, specifically during early and late portions of a given movement. The results for instantaneous knee joint angles and angular velocities were very similar to those of the gold-standard system (mean bias: 0.59–9.52°·s−1), which may be applicable to everyday movement tasks, but may preclude analyses at a clinical level. Although the low-cost sensor platform shows promise an effective monitoring tool, it is not ready yet for a clinical application.
No takes yet. Share an insight, caveat, or question.
Thomson et al. (2026) studied this question.