Answers are grounded in this paper and cite their sources.
A comparative assessment shows significant yield differences in [18F]PSMA-1007 production between synthesis platforms, suggesting implications for clinical use.
Both platforms produced [¹⁸F]PSMA-1007 that consistently met all pharmacopoeial quality control standards. However, the FASTlab2 system demonstrated a statistically significant higher decay-corrected radiochemical yield (43.7% ± 3% vs. 28.5% ± 3.1%, p-value < 0.001 after statistical testing). While this yield difference does not impact radiochemical purity or product safety, it may represent a relevant advantage in terms of production efficiency and available activity for clinical use, which may influence the choice of synthesizer.
No takes yet. Share an insight, caveat, or question.
Cossandi et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: