February 2, 2026Advanced Healthcare Materials2 citationsOpen Access

Real‐Time 3D Ultrasound Imaging with an Ultra‐Sparse, Low Power Architecture

View Full Paper
CMColin MarcusMNMd Osman Goni NayeemASAastha Shah

Key Points

  • To develop a compact, low-power system for effective real-time 3D ultrasound imaging with high resolution.
  • Introduced a convolutional optimally distributed array (CODA) reducing elements from 1024 to 128.
  • Developed a novel chirped data acquisition (cDAQ) architecture for enhanced imaging depth.
  • Implemented a new signal processing methodology for optimal image reconstruction.
  • Achieved deep imaging capabilities exceeding 11 cm in depth.
  • Attained high axial resolution under 600 µm.
  • Demonstrated 29.6x reduction in power consumption and 18 V drive voltage.
  • Validated system performance in vitro and in vivo, successfully detecting tumors and cysts.

Abstract

ABSTRACT Effective resource‐constrained volumetric ultrasound imaging requires compact, low‐power systems capable of wide‐angle real‐time 3D imaging to accommodate small changes in placement by the operator. However, obtaining such images requires an excessive O(N 2 ) channel count, bulky electronics, and high power consumption. We introduce an end‐to‐end system architecture to enable high‐resolution, real‐time 3D ultrasound imaging in a portable form factor. We present: a convolutional optimally distributed array (CODA) geometry that drastically reduces the number of elements (from 1024 to 128), a novel chirped data acquisition (cDAQ) architecture that enhances imaging depth while operating with a 25.3 dB lower transmit amplitude than a pulsed system, and an associated new signal processing methodology. We experimentally demonstrate our system's ability to perform deep (> 11 cm), high axial resolution (< 600 µm), and wide‐angle (57°) imaging, while simultaneously reducing power consumption (29.6x reduction) and drive voltage (18 V). We validated our system in vitro and further performed in vivo human trials, demonstrating the ability to detect both tumors and cysts in breast tissue. This new architectural approach will unlock a new class of medical devices with enhanced diagnostic and long‐term monitoring capabilities and open up future wearable designs of real‐time 3D ultrasound systems.