Finite element analysis compares cement-screw and metal block techniques for tibial defects in total knee arthroplasty, suggesting safety implications for 10 mm defects.
Objective To compare the biomechanical properties of the cement-screw technique with the metal block extension stem technique in reconstructing Anderson Orthopaedic Research Institute (AORI) type 2 tibial defects in total knee arthroplasty using the finite element method, in order to provide a basis for clinical selection. Methods Non-inclusive 5-mm and 10-mm depth AORI type 2 tibial defects were modeled using the finite element method. The cement-screw technique and the metal block extension stem technique were applied for reconstruction, resulting in a total of four sets of finite element models. Each group of models was tested under two loads: horizontal walking and descending stairs. The von Mises stress distributions in the tibia, prosthesis, and bone defect regions, as well as the peaks of micromotion at the prosthesis-tibia interface, were measured. Results In the AORI type 2 tibial defect model, the cement-screw group, when reconstructing 5 mm and 10 mm defects under horizontal walking and descending stairs loads, exhibited higher maximum tibial stress (5 mm: 78.0–140 MPa; 10 mm: 80.9–151 MPa), proximal tibial defect area stress (5 mm: 11.3–25.3 MPa; 10 mm: 10.8–24.1 MPa), and peak micromotion values (5 mm: 9.90–26.99 μm; 10 mm: 11.94–31.98 μm) compared to the metal block extension stem group (tibial stress 5 mm: 73.2–130 MPa, 10 mm: 66.6–118 MPa; defect area stress 5 mm: 7.83–16.3 MPa, 10 mm: 8.54–18.8 MPa; peak micromotion 5 mm: 6.56–14.72 μm, 10 mm: 8.92–24.09 μm). However, prosthesis stresses were lower in the cement-screw group (5 mm: 87.1–183 MPa; 10 mm: 60.2–158 MPa) than in the metal block extension stem group (5 mm: 101–194 MPa, 10 mm: 92.7–167 MPa). Under horizontal walking loading, the two techniques showed no superiority of one over the other in terms of the von Mises stresses and the peaks of micromotion; however, under descending stairs loads, the maximum stress in the tibia of the cement-screw group with a 10-mm defect exceeded 150 MPa (151 MPa), indicating a potential fracture risk, and the peaks of micromotion was smaller in all models. Conclusion The findings of this study indicate that the cement-screw technique is more cost-effective and convenient for repairing 5-mm defects and is appropriate for reconstruction of this size. However, when the bone defect reaches 10 mm, the cement-screw technique may elevate the risk of fracture, and thus, safety concerns must be taken into account. In contrast, the metal block extension stem technique offers a better balance between effectiveness and safety, making it the preferred option for defects of this size.
No takes yet. Share an insight, caveat, or question.
Lu et al. (2026) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: