Key result
HPSD ablation linked to abrupt impedance rise and circumferential catheter thrombus after 11 seconds.
Answers are grounded in this paper and cite their sources.
HPSD ablation using a six-hole irrigated catheter can cause localized overheating and ring-shaped char formation at the anterior wall of the right PVs, highlighting the need for continuous impedance monitoring.
During circumferential ablation of the right pulmonary veins (PVs), high-power, short-duration (HPSD) ablation at the anterior wall of the right PVs, corresponding to the left atrial septal side, resulted in an abrupt impedance increase from 120 to 180 Ω, occurring 11 s after RF delivery initiation (A). After an abrupt impedance rise, withdrawal of the ablation catheter from the long sheath revealed a circumferential thrombus adhering directly above the irrigation holes at the catheter tip (B). The catheter tip after thrombus removal is shown in (C), and the ring-shaped thrombus that adhered to the catheter tip is displayed in (D).
No takes yet. Share an insight, caveat, or question.
Yamauchi et al. (2026) studied this question. High-power, short-duration ablation caused abrupt impedance rise and formation of a circumferential thrombus adhering to irrigation holes on the catheter tip after 11 seconds.