Answers are grounded in this paper and cite their sources.
Validation study assesses AI quantitative coronary angiography software's performance in Europe, indicating strong applicability for clinical decisions.
Artificial intelligence-based quantitative coronary angiography (AI-QCA) has recently emerged as a promising tool for real-time lesion assessment in cardiology. We aimed to validate a novel AI-QCA software, trained on a Korean dataset, in a European cohort. We analyzed 556 lesions from 252 subjects in two European datasets. The AI-QCA system performed automated vessel segmentation and measurements of minimum lumen diameter, proximal and distal reference diameters, percent diameter stenosis (%DS) and lesion length. The performance of AI-QCA was assessed using both automated and manual frame selection methods, with all measurements validated against expert manual QCA. AI-QCA achieved a lesion detection rate of 86.2% in automated frame selection. AI-QCA and manual QCA showed strong agreement (Pearson’s r > 0.90, R2 > 0.8 for all QCA measurements). For %DS categorization (<50%, 50% to <70%, and ≥70%), 433 lesions were classified into the same category by both methods, with a weighted κ of 0.832 (95% CI, 0.743–0.922). Vessel segmentation achieved a mean DSC of 0.953. This study validated the performance of AI-QCA using a European dataset and demonstrated high lesion detection rate and its strong agreement with manual QCA, which supports its applicability for real-time clinical decision-making during percutaneous coronary intervention.
No takes yet. Share an insight, caveat, or question.
Lee et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: