Demonstrates the accurate measurement of quercetin in juices, suggesting a useful method for food analysis.
In this paper, a pencil graphite electrode was pretreated using chronoamperometry technique in phosphate buffer solution (pH=7.0) for sensitive determination of quercetin. Oxidation of quercetin was investigated using pretreated pencil graphite electrode and anodic stripping differential pulse voltammetry. Under optimal conditions, the anodic current of quercetin exhibited linear response to its concentration in the range from 0.001 to 1.5 mu mol/L with the limit of detection of 0.3.10(-3) mu mol/L. The proposed method was successfully applied for the determination of quercetin in cranberry and blackcurrant juices with recovery rate from 93.2 to 94.7 %. Solid-phase extraction was found to be necessary prior to voltammetric determination of quercetin in fruit juice samples using pretreated pencil graphite electrode.
No takes yet. Share an insight, caveat, or question.
Dilgin et al. (2015) studied this question.