This analysis calibrates time signals from scintillation counters, highlighting efficiency and resolution in detector performance.
The procedure to calibrate in time the scintillation counters of the HARP TOF-WALL using cosmic muons is described in detail. The analysis of the collected cosmic events was also used to evaluate the performance of the detectors in terms of efficiency and time resolution. All scintillation counters resulted to be fully efficient. The procedure allowed to precisely align in time the signals of all counters so that an overall intrinsic time resolution of 9 160 ps could be achieved