Demonstrates high-precision detection of overlapping ammonia spectral lines, suggesting improved environmental monitoring techniques.
Ammonia (NH₃) emissions pose significant hazards to both the environment and human health, making high‐precision monitoring of ambient NH₃ concentrations particularly crucial. In this study, a high‐precision NH₃ detection system was designed based on near‐infrared dual‐comb spectroscopy (DCS). A laser source with a central wavelength of 1531 nm is employed as the laser relay; by locking the relative frequencies of two optical combs to the relay laser separately, dual‐comb with a central frequency of 1531 nm, a repetition rate of 518.25 Hz, and an output power of 30 μW is generated. A multi‐pass gas absorption cell with a base length of 0.113 m and 250 reflections is utilized to achieve an effective optical path length of approximately 25 m. Under conditions of 296.15 K and 1 atm, three overlapping absorption lines at 6528.78 cm⁻¹, 6528.88 cm⁻¹, and 6529.18 cm⁻¹ of 1000 ppm NH₃ are detected, verifying the feasibility of high‐resolution dual‐comb spectroscopy for measuring overlapping spectral lines. Allan variance analysis is performed on the experimental data, yielding a minimum detection limit (MDL) of 0.00128 for absorbance at an integration time of 972 s, corresponding to a concentration MDL of ~2 ppm. This research provides technical insights for achieving high‐precision measurement and analysis of overlapping spectral lines.
No takes yet. Share an insight, caveat, or question.
Rong et al. (2026) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: