Research analyzes a grating ruler system achieving high precision in micro-displacement measurements, suggesting broad application prospects.
To achieve high-precision measurement of micro-displacement, this study takes the Moiré fringe principle as the core and designs and fabricates a modular grating ruler measurement system. The system comprises a light source, a scale grating and an index grating with grating pitches of 1.4mm and 1.6mm respectively, a low-speed motor traction device, a photoresistor sensor, and a 51 single-chip microcomputer (SCM) calculation unit.The experiment adopts the point-by-point measurement method, with each group of measurements repeated 10 times and the average value taken. The results show that the system has an effective measurement range of 120mm, a measurement resolution of 1μm, a repeat positioning accuracy of ±0.5μm, and an average relative error of 7.3%.Boasting advantages such as modular design, easy operation, and controllable cost, the device can provide a reliable solution for micro-displacement measurement in fields like precision manufacturing and optical instruments, and possesses broad application prospects.
No takes yet. Share an insight, caveat, or question.
Chuanyu et al. (2025) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: