Answers are grounded in this paper and cite their sources.
This work demonstrates improved bandwidth and gain in a dual-band patch antenna for skin implants, suggesting its practical application potential.
This work presents a compact dual‐band patch antenna designed for skin implant applications. The antenna features a robot‐shaped patch structure on FR‐4 substrate and resonates at dual bands of 915 and 2430 MHz. Three techniques are employed to miniaturize the antenna: a shorting pin between the patch and ground, defected ground structure (DGS), and utilization of tissue electrical properties. Thus, the optimized antenna volume is found to be 8 × 12 × 1.36 mm 3 , with near‐perfect impedance matching. The antenna also achieves enhanced impedance bandwidths of 15.74% and 15.64% and gains of −29 and −16.6 dBi at 915 and 2430 MHz, respectively. Its specific absorption rates (SARs) are found to be acceptable, with values of 448 and 221 W/kg in the 1 g tissue standard at the respective bands. Furthermore, the antenna can communicate with the receiver antenna at a high data rate of 1 Mbps over 8–12 m, where the distance can be extended to 25–40 m by choosing a low data rate of 100 Kbps. The antenna is measured for its return loss ex vivo using chicken tissue, which is found to be in close agreement with the simulation results. Thus, the balanced performance in terms of compact size, large bandwidth, and comparable gain and SAR makes the proposed antenna a promising candidate for practical applications.
No takes yet. Share an insight, caveat, or question.
Mainul et al. (2026) studied this question.