This article details the upgraded ptychography endstation's performance in nanoscale imaging, highlighting its significance.
Following our initial demonstration of ptychography at the Taiwan Photon Source (TPS), the system has been substantially upgraded. This article details the design and performance of the new hard X-ray 2D ptychography endstation, TPS 25A2. A zone-plate focusing system has been inserted downstream of the previous Kirkpatrick-Baez mirror optics at the TPS 25A2 beamline. Operating at a photon energy of 8.979 keV, the endstation routinely achieves a spatial resolution of sub-10 nm in 2D, with a highest achieved resolution of 6.9 nm on a standard Siemens star test pattern. We present the full instrumental specifications and performance benchmarks, and showcase the endstation's capabilities with imaging examples of inorganic nanoparticles and a complex biological structure. The establishment of TPS 25A2 provides the scientific community with a powerful and accessible tool for nanoscale imaging.