Demonstrates a navigation guide's efficacy for inserting cortical bone trajectory screws in formerly instrumented vertebrae, suggesting enhanced surgical precision.
Purpose Traditional surgical interventions for adjacent segment disease (ASD) repair commonly mandate the exposure of all pre-existing hardware. Cortical bone trajectory (CBT) screws have been innovated to amplify the contact surface between the screw and cortical bone, offering a promising approach to mitigate the challenge sassociated with traditional ASD repair surgeries. Methods In this study, a tangible manifestation of the cortical bone trajectory screw navigation guide was produced by utilizing a 3D printer to facilitate the insertion of cortical bone trajectory screws. A comprehensive set of radiographic images and computed tomography scans was gathered to evaluate the accuracy of K-wire through CBT screw guide inside the vertebrae which was previously implanted with a pedicle screw. Results The variance in the K-pin circle centre distance before and after surgery ranges from 0.2 mm to 3.54 mm. The K-pin angle difference falls within the range of 0.4° to 5.8° in the sagittal plane, ranges from 1.46° to 1.7° in the transverse plane. Conclusion This research process has successfully yielded a navigation guide that holds promise for inserting CBT screw even with previously pedicle screw at the same vertebrae.
No takes yet. Share an insight, caveat, or question.
Tai et al. (2026) studied this question.