Key result
TEE identifies CT thrombus mimic post-WATCHMAN FLX, enabling safe anticoagulation withdrawal.
Why the study?
Device-related thrombus remains a complication after left atrial appendage closure, and imaging modalities can have diagnostic pitfalls.
Does adjunctive transesophageal echocardiography clarify thrombus location and prevent unnecessary oral anticoagulation in post-LAAC patients with atypical cardiac CT findings suggesting device-related thrombus?
Case Report (n=1)
No
Does adjunctive transesophageal echocardiography clarify thrombus location and prevent unnecessary oral anticoagulation in post-LAAC patients with atypical cardiac CT findings suggesting device-related thrombus?
In patients after left atrial appendage closure with atypical cardiac CT findings meeting criteria for definite device-related thrombus, adjunctive transesophageal echocardiography can identify fabric-confined thrombus mimics and prevent unnecessary long-term oral anticoagulation.
Answers are grounded in this paper and cite their sources.
TEE may distinguish fabric mimics from true DRT on post-LAAC CT, avoiding unnecessary anticoagulation; leaves open need for prospective validation.
BACKGROUND: Left atrial appendage closure (LAAC) is an established stroke-prevention option for atrial fibrillation patients unsuitable for long-term oral anticoagulation (OAC), while device-related thrombus (DRT) remains a complication. Cardiac computed tomography (CT) and transesophageal echocardiography (TEE) are modalities for its detection. CASE SUMMARY: A patient after LAAC using WATCHMAN-FLX (Boston Scientific) underwent routine cardiac CT, which demonstrated an irregular, hypoattenuated surface lesion fulfilling criteria for definite DRT. However, the mass had a somewhat atypical configuration. Additional TEE demonstrated bulging confined to the device fabric with no thrombus extending into the left atrium. OAC was not intensified, and the patient remained stable. DISCUSSION: This case illustrates a pitfall of CT-based DRT diagnosis in WATCHMAN devices. An eccentric fabric elevation can mimic a protruding thrombus. TAKE-HOME MESSAGE: In post-LAAC patients with atypical CT findings suggesting DRT, adjunctive TEE is helpful to clarify thrombus location and prevent unnecessary OAC.
No takes yet. Share an insight, caveat, or question.
Nagai et al. (2026) conducted a case report in Suspected device-related thrombus after left atrial appendage closure with a WATCHMAN FLX device (n=1). Transesophageal echocardiography (TEE) evaluation of CT-suspected device-related thrombus vs. Routine cardiac computed tomography (CT) was evaluated on Clarification of device-related thrombus vs fabric-confined mimic and subsequent thromboembolic/bleeding events. Transesophageal echocardiography clarified a 6.9-mm CT-detected lesion as a fabric-confined thrombus mimic after WATCHMAN FLX closure, enabling safe anticoagulation withdrawal.