Retrospective comparison reveals 40-keV virtual monoenergetic imaging enhances lesion visibility in breast cancer, suggesting potential as an accessible alternative to DCE-MRI.
Purpose This study investigated whether contrast enhanced whole-body photon-counting detector (PCD)-CT performed for staging in the prone position can provide breast lesion conspicuity comparable to dynamic contrast-enhanced T1-weighted imaging (DCE-T1WI). Methods This single-center, retrospective diagnostic feasibility study included 22 women with surgically confirmed primary breast cancer who underwent preoperative prone PCD-CT and DCE-MRI. Then, 40- and 70-keV virtual monoenergetic image (VMI) from a single arterial-phase PCD-CT acquisition were reconstructed. Three readers independently evaluated 40- and 70-keV VMI and DCE-T1WI. The primary endpoint was lesion conspicuity. The secondary endpoints included, size error, detection rate, background parenchymal enhancement (BPE), subjective noise, and reading time. Results Lesion conspicuity was highest on 40-keV VMI, which outperformed 70-keV VMI and DCE-T1WI. Absolute size error was numerically larger for 40-keV VMI than for DCE-T1WI, but this difference was not statistically significant. Directional error indicated that 40-keV VMI significantly underestimated lesion size relative to DCE-T1WI. Detection rates showed an overall difference among modalities, but pairwise differences were not statistically significant. Further, 40-keV VMI had a higher BPE and a greater perceived noise than 70-keV VMI, and it required a longer reading time than DCE-T1WI. The radiation dose did not differ between VMI energy levels as both were reconstructed from the same acquisition. Conclusion Arterial-phase 40-keV VMIs from prone PCD-CT demonstrated the highest lesion conspicuity and showed better detection capability than 70-keV VMI. In addition, their performance was comparable to or superior to DCE-T1WI. This finding was achieved despite trade-offs in increased noise, BPE and size underestimation. Thus, 40-keV VMI may serve as a useful complementary modality for local breast assessment and may be considered a potential alternative for lesion localization when MRI is not feasible, despite potential underestimation of lesion extent.
No takes yet. Share an insight, caveat, or question.
Kawasaki et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: