Demonstrates reduced oxidative stress in hemodialysis patients using a new dialysate reduction system, suggesting potential clinical benefits.
〔Objective〕 In hemodialysis patients, oxidative stress can be exacerbated by the dialysis procedure itself, bioincompatibilities with the dialysis apparatus, or by the loss of antioxidants during dialysis. Here, we report the development of a novel dialysate reduction system and estimate its effect on oxidative stress. 〔Methods〕 A total of 14 patients receiving maintenance hemodialysis were enrolled in this study. We performed hemodialysis for 4 weeks using a system to dissolve hydrogen gas in an acid concentrate. We then estimated the changes in the oxidation reduction potential (ORP) and biochemical data. 〔Results〕 In the conventional dialysis system, the ORP of the blood had increased by 68.1 ± 30.5 mV at the exit of the dialyzer, compared with the value at the entrance, whereas the use of the reduction device suppressed the increase to 29.3 ± 20.5 mV (P = 0.012). The ferritin level also decreased from 91.9 ± 77.0 ng/mL to 63.0 ± 46.7 ng/mL (P = 0.035). 〔Conclusion〕 A novel dialysate reduction system suppressed elevations in the ORP of blood. The ferritin level, which is an indicator of chronic inflammation, was also reduced. This system might be useful for reducing oxidative stress in hemodialysis patients.
No takes yet. Share an insight, caveat, or question.
Otsubo et al. (2018) studied this question.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: