Why the study?
Does the choice of material model impact the predicted stress pattern and coaptation parameters in patient-specific simulations of stentless aortic valve implantation?
Does the choice of material model impact the predicted stress pattern and coaptation parameters in patient-specific simulations of stentless aortic valve implantation?
Computational simulations demonstrate that the choice of material model significantly impacts the predicted post-operative performance of stentless aortic valve implants.
Material model choice may alter simulated stentless valve outcomes; leaves open standardization before clinical planning use.
In some cases of aortic valve leaflet disease, the implant of a stentless biological prosthesis represents an excellent option for aortic valve replacement (AVR). In particular, if compared with the implant of mechanical valves, it provides a more physiological haemodynamic performance and a reduced thrombogeneticity, avoiding the use of anticoagulants. The clinical outcomes of AVR are strongly dependent on an appropriate choice of both prosthesis size and replacement technique, which is, at present, strictly related to surgeon's experience and skill. This represents the motivation for patient-specific finite element analysis able to virtually reproduce stentless valve implantation. With the aim of performing reliable patient-specific simulations, we remark that, on the one hand, it is not well established in the literature whether bioprosthetic leaflet tissue is isotropic or anisotropic; on the other hand, it is of fundamental importance to incorporate an accurate material model to realistically predict post-operative performance. Within this framework, using a novel computational methodology to simulate stentless valve implantation, we test the impact of using different material models on both the stress pattern and post-operative coaptation parameters (i.e. coaptation area, length and height). As expected, the simulation results suggest that the material properties of the valve leaflets affect significantly the post-operative prosthesis performance.
No takes yet. Share an insight, caveat, or question.
Auricchio et al. (2012) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: