Answers are grounded in this paper and cite their sources.
When irradiated in cylindrical, rectangular, and cubic phantoms, nanoDots show a maximum angular dependence of 1% or less at an incidence angle of 90°. For a sample of 78 new nanoDots, the range of their relative intrinsic sensitivity is 0.92-1.09. For a sample of ten nanoDots, on average, the mass in the sensitive layer is 73.1% Al2O3:C and 26.9% polyester. The mass density of the sensitive layer of a nanoDot disc is between 2.42 and 2.58 g/cm(3). The angular dependence is not related to Al2O3:C loading of the nanoDot disc. The nanoDot at the depth of maximum dose has no more angular dependence than a parallel-plate ion chamber.
No takes yet. Share an insight, caveat, or question.
Paul A. Jursinic (2015) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: