The gain of a slow-acting compression hearing aid should place the output within 5 dB of MCL. The output for low and medium inputs should approximate MCL and the output for high input levels should be slightly above MCL. This pattern of gain may be obtained with mild compression ratios and a gain rule that places a speech input of 70 dB at MCL.
No takes yet. Share an insight, caveat, or question.
Neuman et al. (1995) studied this question.