Answers are grounded in this paper and cite their sources.
The connection between Johnson noise and blackbody radiation is discussed, using a simple thermodynamic model. A microwave radiometer is described together with its theory of operation. The experimentally measured root mean square fluctuation of the output meter of a microwave radiometer (0.4°C) compares favorably with a theoretical value of 0.46°C. With an r-f band width of 16 mc/sec., the 0.4°C corresponds to a minimum detectable power of 10−16 watt. The method of calibrating using a variable temperature resistive load is described.
No takes yet. Share an insight, caveat, or question.
R. H. Dicke (1946) studied this question.