Answers are grounded in this paper and cite their sources.
Flow diversion is a promising technique for cerebral aneurysm treatment but presents new and sometimes unique challenges. Transient reversible narrowing due to device reconfiguration of the pipeline embolization device (PED) has not been previously described. Here we describe a patient with dolichoectasia of the distal right internal carotid artery with an associated saccular sidewall aneurysm who developed asymptomatic circumferential narrowing of the proximal aspect of the PED in the first month following deployment. The patient was followed conservatively and demonstrated partial resolution of device narrowing on 6 month follow-up imaging.
No takes yet. Share an insight, caveat, or question.
Miller et al. (2014) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: