Common electrochemical aptamer-based (E-AB) biosensors employing methylene blue (MB) redox reporters suffer from significant pH-induced signal variations. Consequently, dynamic pH fluctuations in biofluids, such as in sweat, can greatly distort the measured signal and lead to inaccurate target quantification. Here, we introduce a self-correction strategy that enables accurate analyte quantification by compensating for pH-induced signal variations in MB-based aptasensors using the Nernstian shift of the MB peak potential. The highly reproducible reversible pH dependence of the MB peak potential enables real-time monitoring of the sample pH that provides a continuously synchronized built-in compensation of the pH interference, thereby eliminating the need for external pH sensors or additional measurement steps. This precise yet simple and effective self-correction of pH effects ensures that the Square Wave Voltammetry (SWV) signal reliably reflects real-time analyte variations, as demonstrated using a cortisol MB-labeled E-AB biosensor in both in vitro and on-body settings. In vitro measurements in artificial sweat across pH 5.5-7.5 showed excellent correlation with real-time pH and cortisol changes, confirming the reliability of the peak potential-based correction. On-body measurements using an epidermal wearable patch showed cortisol changes in sweat that would have been missed without this signal-correction method. This approach is broadly applicable to other MB-labeled E-AB biosensors and biofluids, providing a robust strategy for continuous on-body monitoring.
Building similarity graph...
Analyzing shared references across papers
Loading...
Barak Sabbagh
University of California, San Diego
Muhammad Inam Khan
Jinnah Postgraduate Medical Center
G.C. Kim
University of California, San Diego
ACS Sensors
University of California, San Diego
Building similarity graph...
Analyzing shared references across papers
Loading...
Sabbagh et al. (Tue,) studied this question.
synapsesocial.com/papers/69d894526c1944d70ce0533a — DOI: https://doi.org/10.1021/acssensors.6c00737
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: