Experimental evaluation demonstrates 3D in-system positioning calibration in a proof-of-concept scanner, indicating feasible calibration for complex light-sharing PET detector architectures.
The functionality of the 3D in-system positioning calibration method was successfully demonstrated with the measurements within a PoC scanner configuration with similar positioning performance as the bench-top fan-beam setup. The in-system calibration method can be used to calibrate and test fully assembled PET systems to enable more complex light-sharing detector architectures in, for example, large PET systems with many detectors. The acquired data can further be used for more complex energy and time calibrations.