Laser photoacoustic spectroscopy (LPAS) is highly suitable for the detection of ethene in air due to the overlap between its strongest absorption lines and the wavelengths accessible by high-powered CO2 lasers. Here, we test the ability of LPAS to measure ethene in ambient air by comparing the measurements in urban air with those from a gas chromatography flame-ionization detection (GC-FID) instrument. Over the course of several days, we obtained quantitative agreement between the two measurements. Over this period, the LPAS instrument had a positive offset of 330 +/- 140 pptv (parts-per-trillion by volume) relative to the GC-FID instrument, possibly caused by interference from other species. The detection limit of the LPAS instrument is currently estimated around 1 ppbv and is limited by this offset and the statistical noise in the data. We conclude that LPAS has the potential to provide fast-response measurements of ethene in the atmosphere, with significant advantages over existing techniques when measuring from moving platforms and in the vicinity of emission sources.
No takes yet. Share an insight, caveat, or question.
Kuster et al. (2005) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: