Answers are grounded in this paper and cite their sources.
We studied the reproducibility of X-ray dose rate measurements using radioluminescent (RL) single-mode (SM) and multi-mode (MM) N-doped silica-based optical fibers (OFs) in the [22 mGy(SiO2)/s–6.6 Gy(SiO2)/s] range. The influence of the probe manufacturing parameters on the dispersion of the RL sensitivity coefficient ( <tex-math notation="LaTeX">χ </tex-math> in counts <tex-math notation="LaTeX">· </tex-math> mGy <tex-math notation="LaTeX">⁻¹· s </tex-math> ) is discussed for 20 manufactured probes. In particular, we demonstrated how the quality control of the splice between the N-doped sample and its transport fiber can efficiently reduce the dispersion of those coefficients. In addition, also the impact of measuring the RL fiber length was studied. Thanks to this study, we could estimate the achievable performances of this potential fiber-based dosimeter, avoiding the individual calibration procedure of each probe. The good performances in terms of <tex-math notation="LaTeX">χ </tex-math> dispersion (17% for the MM optical and 3.8% for the SM OFs) demonstrate the high potential of this technology for implementation in harsh environments.
No takes yet. Share an insight, caveat, or question.
Fricano et al. (2023) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: