Answers are grounded in this paper and cite their sources.
We investigated the potential of a multimode (50-μm core diameter) nitrogen doped silica-based optical fiber for X-rays, protons, and y-rays radiation detection and dosimetry. X-rays results confirm that this N-doped fiber presents a strong radioluminescence (RL) around 550 nm when exposed to radiation. The RL observed from a few centimeter long samples linearly depend on the dose rate at least in the 1-mGy/s-50-Gy/s range. Low dose rate y-ray t ests demonstrate the radiation detection at dose rate as low as 40 μGy/s with a 2-m-long fiber sample. This fiber is also associated with an optically stimulated luminescence (OSL) when exposed to a 1064-nm laser light postirradiation. This OSL signature provides shortly after the irradiation a precise estimation of the accumulated dose at least in the 1-Gy-10-kGy range. Under proton exposure, we benchmarked the performances of the N-doped fiber with the ones of a conventional Markus chamber used by the TRIUMF proton-therapy team. This comparison highlights the potential of the small-size N-doped fiber to partially overcome the known limitations of commercial fiber-based dosimeters for medical applications. The obtained results are very promising for the future design of more complex spatially resolved beam monitoring systems.
No takes yet. Share an insight, caveat, or question.
Girard et al. (2018) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: