Answers are grounded in this paper and cite their sources.
We have analyzed the possibility of using the resistive coaxial cable, Thermocoax(R) Philips as a microwave cryofilter for single electron experiments performed in a top-loading system dilution refrigerator. The biasing and signal lines made of this cable are assumed to link a single electron device kept at the lowest temperature Td to the rf filter having a temperature Tf of several kelvin. The attenuation in a wide frequency range has been calculated and it turned out that a 40-cm-long piece of this cable with an outer diameter of 0.5 mm could drastically reduce the effect of noise from the 50 Ω source anchored at Tf (below 25 K), making its effect comparable with the effect of the thermal fluctuations in the device for Td≊40 mK. The calculated attenuation is in reasonable agreement with the data measured at frequencies up to 18 GHz.
No takes yet. Share an insight, caveat, or question.
A. B. Zorin (1995) studied this question.