Why the study?
The Impella device is routinely used during coronary interventions, but device-related iatrogenic injury requiring surgical repair is rarely reported.
Highlights a rare but severe complication of Impella device insertion (iatrogenic aortic valve injury), suggesting the need for follow-up echocardiography after device removal.
Answers are grounded in this paper and cite their sources.
Delayed aortic insufficiency post-Impella may warrant echocardiographic surveillance; extends known complications but leaves incidence and prevention open.
The Impella device is used routinely during coronary interventions, and device-related iatrogenic injury requiring surgical repair is rarely reported. We report a case of iatrogenic aortic valve injury due to Impella device insertion, necessitating aortic valve replacement surgery. A 55-year-old woman with acute anteroseptal myocardial infarction underwent a successful percutaneous coronary intervention and Impella insertion for hemodynamic support. Four months later, she experienced congestive heart failure; transthoracic echocardiography revealed new-onset severe aortic insufficiency. Aortic valve replacement was performed without complications. Follow-up transthoracic echocardiography should be performed after Impella removal to avoid delay in treating residual aortic insufficiency.
No takes yet. Share an insight, caveat, or question.
Saitoh et al. (2020) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: