Systematic review of literature following release of the 2013 guideline reaffirms the use of a validation set of at least 60 cases, establishing intraobserver diagnostic concordance between WSI and glass slides and the use of a 2-week washout period between modalities. Although all discordances between WSI and glass slide diagnoses discovered during validation need to be reconciled, laboratories should be particularly concerned if their overall WSI-glass slide concordance is less than 95%.