Answers are grounded in this paper and cite their sources.
The horizontal displacement in all groups was less than 10 µm. With the hand tightening torque, a high vertical displacement of over 100 µm occurred in PEEK scan bodies (Myfit and Dentium). Therefore, it is recommended to apply a tightening torque of 5 Ncm instead of a hand tightening torque.
No takes yet. Share an insight, caveat, or question.
Kim et al. (2020) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: