OSLDs can be successfully used for HDR in vivo dosimetry. However, for the measurements to be meaningful one must account for the angular dependence, volume-averaging, and the greater sensitivity to Ir-192 gamma rays than to 6 MV x-rays if 6 MV x-rays were used for OSLD calibration. The limitations of the treatment planning algorithm must be understood, especially for surface dose measurements. Use of in vivo dosimetry for HDR brachytherapy treatments is feasible and has the potential to detect and prevent gross errors. In vivo HDR brachytherapy should be included as part of the QA for a HDR brachytherapy program.
No takes yet. Share an insight, caveat, or question.
Sharma et al. (2013) studied this question.