Key result
Consumer wearables match 12-lead ECGs for detecting AF, achieving 100% sensitivity and high specificity.
Why the study?
Despite numerous individual validation studies, a direct comparison of interdevice performance of consumer wearable devices for atrial fibrillation detection was lacking.
Do consumer wearable devices accurately detect atrial fibrillation compared to a 12-lead ECG in cardiology outpatients?
Cross-Sectional (n=122)
Random sequence
No
Do consumer wearable devices accurately detect atrial fibrillation compared to a 12-lead ECG in cardiology outpatients?
Answers are grounded in this paper and cite their sources.
Consumer wearable devices using ECG or PPG are highly accurate in discriminating between sinus rhythm and atrial fibrillation, with comparable performance across different devices.
Background: Consumer-oriented wearable devices (CWDs) such as smartphones and smartwatches have gained prominence for their ability to detect atrial fibrillation (AF) through proprietary algorithms using electrocardiography or photoplethysmography (PPG)-based digital recordings. Despite numerous individual validation studies, a direct comparison of interdevice performance is lacking. Objective: This study aimed to evaluate and compare the ability of CWDs to distinguish between sinus rhythm and AF. Methods: Patients exhibiting sinus rhythm or AF were enrolled through a cardiology outpatient clinic. The participants were instructed to perform heart rhythm measurements using a handheld 6-lead electrocardiogram (ECG) device (KardiaMobile 6L), a smartwatch-derived single-lead ECG (Apple Watch), and two PPG-based smartphone apps (FibriCheck and Preventicus) in a random sequence, with simultaneous 12-lead reference ECG as the gold standard. Results: A total of 122 participants were included in the study: median age 69 (IQR 61-77) years, 63.9% (n=78) men, 25% (n=30) with AF, 9.8% (n=12) without prior smartphone experience, and 73% (n=89) without experience in using a smartwatch. The sensitivity to detect AF was 100% for all devices. The specificity to detect sinus rhythm was 96.4% (95% CI 89.5%-98.8%) for KardiaMobile 6L, 97.8% (95% CI 91.6%-99.5%) for Apple Watch, 98.9% (95% CI 92.5%-99.8%) for FibriCheck, and 97.8% (95% CI 91.5%-99.4%) for Preventicus (P=.50). Insufficient quality measurements were observed in 10.7% (95% CI 6.3%-17.5%) of cases for both KardiaMobile 6L and Apple Watch, 7.4% (95% CI 3.9%-13.6%) for FibriCheck, and 14.8% (95% CI 9.5%-22.2%) for Preventicus (P=.21). Participants preferred Apple Watch over the other devices to monitor their heart rhythm. Conclusions: In this study population, the discrimination between sinus rhythm and AF using CWDs based on ECG or PPG was highly accurate, with no significant variations in performance across the examined devices.
No takes yet. Share an insight, caveat, or question.
Wouters et al. (2024) conducted a cross-sectional in Atrial Fibrillation (n=122). Consumer wearable devices (KardiaMobile 6L, Apple Watch, FibriCheck, Preventicus) vs. 12-lead reference ECG was evaluated on Sensitivity to detect AF and specificity to detect sinus rhythm (p=.50). Consumer wearable devices demonstrated 100% sensitivity and 96.4% to 98.9% specificity for detecting atrial fibrillation compared to a 12-lead reference ECG, with no significant differences.
Synapse has enriched 10 related papers. Consider them for comparative context: