Why the study?
Pacing-threshold elevation requiring percutaneous retrieval and replacement can complicate leadless pacemaker use, but whether acute elevations can resolve without intervention is unclear.
Loss of capture and high impedance following leadless pacemaker implantation may improve spontaneously during short-term follow-up, potentially avoiding unnecessary invasive device replacement.
Answers are grounded in this paper and cite their sources.
May support short-term observation before revising early high-impedance leadless pacemakers; leaves open incidence and predictors.
Leadless pacemakers (LPMs) offer an advantage over traditional pacemakers that require creation of a generator pocket and connection of the transvenous leads, which may lead to complications. However, device dislodgement, cardiac perforation, and pacing-threshold elevation requiring percutaneous retrieval and device replacement have been reported as serious adverse events occuring with use of LPMs.1Reddy V.Y. Exner D.V. Cantillon D.J. et al.Percutaneous implantation of an entirely intracardiac leadless pacemaker.N Engl J Med. 2015; 373: 1125-1135Crossref PubMed Scopus (367) Google Scholar Herein, we report the case of an LPM-implanted patient who experienced a transient and acute elevation of impedance and pacing threshold. These symptoms improved without requiring device replacement. An 85-year-old man without structural heart disease underwent implantation of an LPM (Micra AV, Medtronic, Minneapolis, MN) for a complete atrioventricular block. The procedure was performed using a previously described technique.2Pachón M. Puchol A. Akerström F. et al.Implantation of the Micra transcatheter pacing system: initial experience in a single Spanish center.Rev Esp Cardiol. 2016; 69: 346-349Crossref PubMed Scopus (13) Google Scholar After insertion of the femoral venous catheter, a 3000-U bolus of intravenous heparin was administered. The pacemaker delivery catheter was directed toward the right ventricular septum, using the gooseneck-shape technique, and deployed in the right ventricular septum. The time until device deployment was slightly longer than usual, because the tip of the delivery catheter slipped easily and could not attach sufficiently to the interventricular septum. We performed a pull-and-hold test and confirmed that the 2 tines were engaged tissue. The measured electrical parameters were appropriate, with a sensed R wave of 8.0 mV, a pacing impedance of 520 Ω, and a pacing threshold of 1.63 V at 0.24 ms. Changes in the electrical parameters are shown in Table 1. The patient was observed for 3 days, with no complications. After 3 days, loss of capture was observed on electrocardiographic monitoring. The measured electrical parameters were a sensed R wave of 3.0 mV, a pacing impedance of > 2500 Ω, and a pacing threshold of 4.61 V at 1.0 ms. The electrical parameters did not change with the patient in sitting and lying positions. Automatic electrode impedance measurement on the third day after LPM implantation was about 500 Ω. Chest radiography and fluoroscopic imaging showed stable device location in the right ventricle, similar to the intraprocedural location (Fig. 1). Fluoroscopic imaging also showed the swinging movement of the LPM. Reimplantation of the LPM was scheduled. However, the next day, manually measured electrical parameters showed a sensed R wave of 3.5 mV, pacing impedance of 400 Ω, and a pacing threshold of 1.5 V at 0.4 ms. Automatic electrode impedance measurement at 2:30 AM on day 4 after LPM implantation was about 400 Ω, which was an acceptable level of impedance. The patient did not receive any oral anticoagulation therapy. We monitored the electrical parameters remotely without any further complications arising.Table 1The timeline of the patient’s electrical parameter measurementsTime and eventElectrical parametersSensed R wave, mVPacing impedance, ΩPacing threshold,V at (ms)Leadless pacemaker implantation (Day 0)8.05201.63 (0.24)Day 3, 2:30 AM500∗Approximate values from automatic electrode impedance measurement.Day 3, 8 AMObserved loss of capture3.0> 25004.61 (1.0)Day 3, 2:30 PM3.021504.61 (1.0)Day 4, 2:30 AM400∗Approximate values from automatic electrode impedance measurement.Day 4, 9 AM3.54001.5 (0.4)∗ Approximate values from automatic electrode impedance measurement. Open table in a new tab In this case, we observed transient high impedance and high threshold in the implanted LPM. A previous study reported elevated pacing thresholds requiring device retrieval and reimplantation in 1.3% of patients with LPMs.1Reddy V.Y. Exner D.V. Cantillon D.J. et al.Percutaneous implantation of an entirely intracardiac leadless pacemaker.N Engl J Med. 2015; 373: 1125-1135Crossref PubMed Scopus (367) Google Scholar This report is the first showing that improvement in the elevated pacing threshold and impedance can occur without an invasive procedure. Implantable cardioverter-defibrillator shock impedance has been reported to gradually rise due to encapsulation fibrosis.3Monkhouse C. Cambridge A. Chow A.W.C. Behar J.M. High-voltage impedance rise; mechanism and management in patients with transvenous implantable cardioverter-defibrillators: a case series.Eur Heart J Case Rep. 2019; 3: 1-8Crossref PubMed Scopus (3) Google Scholar In a study of patients requiring LPM retrieval and replacement, encapsulation and adhesion of LPM were considered contributing factors for elevated impedance.4Minami K. Neužil P. Petrů J. et al.Retrieval of long-term implanted leadless pacemakers: a single-center experience.JACC Clin Electrophysiol. 2020; 6: 1744-1751Crossref PubMed Scopus (7) Google Scholar However, a swinging movement was confirmed in these cases, as not enough time had elapsed since the implantation for adhesion and encapsulation to occur. In the previous report, high pacing impedance and absence of pacing capture were observed during LPM implantation. The transfemoral catheter delivery system was removed, and a thrombus coating the entire distal portion of the LPM, including the cathode, was observed.5Arias M.A. Rubio M.A. Miguel R. Pachón M. Thrombus formation at the tip of a leadless pacemaker causing multiple unnecessary repositioning.Heart Rhythm. 2016; 13: 2265Abstract Full Text Full Text PDF PubMed Scopus (6) Google Scholar In our patient, a small thrombus could have formed between the LPM and the myocardium, temporarily elevating impedance and the pacing threshold before disappearing the next day. Nitinol tines tend to burrow into the myocardial tissue, but the Micra AV is designed with a slight distance between the nitinol tines and the electrodes. Therefore, we assumed that a small gap was present between the electrode and the endocardial surface, in which a clot could form, or into which a foreign object could enter. The time to deploy the device was slightly longer than usual, which may have caused myocardial changes, including inflammation and edema, making it more likely to form a thrombus. A possible reason the sensing did not improve is that the sensing vector changed when the contact between the electrode and the endocardial surface slightly changed after the disappearance of a clot or foreign object. However, from this single case only, we could not determine whether high impedance and thrombus formation were related, and the mechanism remains a matter for speculation; further investigations are needed. If an LPM that has been stably pacing for several days shows a sudden loss of capture and high impedance, it may improve with a few days of follow-up, as seen in this case.Novel Teaching Points•Loss of capture and high impedance following LPM implantation may improve during follow-up, owing to the disappearance of a small temporary thrombus that had formed between the LPM and the myocardium. •Loss of capture and high impedance following LPM implantation may improve during follow-up, owing to the disappearance of a small temporary thrombus that had formed between the LPM and the myocardium. The research reported has adhered to relevant ethical guidelines. The authors have no funding sources to declare.
No takes yet. Share an insight, caveat, or question.
Hagiwara et al. (2022) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: