July 11, 2022Frontiers in Cardiovascular Medicine56 citationsOpen Access

Smartwatch-Based Blood Pressure Measurement Demonstrates Insufficient Accuracy

MFMaarten FalterMSMartijn ScherrenbergKDKaren Driesen

Structured PICO

Does smartwatch-based blood pressure measurement provide accurate readings in normotensive and hypertensive patients compared to traditional sphygmomanometers?

P
Population
Normotensive and hypertensive patients
I
Intervention
Smartwatch-based blood pressure measurement (Samsung Galaxy Watch Active 2)
C
Comparator
Traditional non-invasive sphygmomanometers (reference standard)
O
Outcome
Accuracy of blood pressure measurement (bias toward calibration point)surrogate

Smartwatch-based blood pressure measurement using the Samsung Galaxy Watch Active 2 is not yet ready for clinical use due to systematic bias and insufficient accuracy.

Abstract

This study demonstrates that the BP measurements by the Samsung Galaxy Watch Active 2 show a systematic bias toward a calibration point, overestimating low BPs and underestimating high BPs, when investigated in both normotensive and hypertensive patients. Standards for traditional non-invasive sphygmomanometers are not met, but these standards are not fully applicable to cuffless devices, emphasizing the urgent need for new standards for cuffless devices. The smartwatch-based BP measurement is not yet ready for clinical usage. Future studies are needed to further validate wearable devices, and also to demonstrate new possibilities of non-invasive, high-frequency BP monitoring.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Falter et al. (2022) studied this question.

synapsesocial.com/papers/69d9d8f70f32475823a3c877https://doi.org/10.3389/fcvm.2022.958212
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Measuring agreement in method comparison studies1999 · 8,847 citations
  2. 22020 International Society of Hypertension Global Hypertension Practice Guidelines2020 · 4,219 citations
  3. 3Cuffless blood pressure monitoring from a wristband with calibration-free algorithms for sensing location based on bio-impedance sensor array and autoencoder2022 · 80 citations
  4. 4A reliable index for the prognostic significance of blood pressure variability2005 · 537 citations
  5. 5Validation of an Automated and Adjustable Blood Pressure System for Use with a Public Health Station2020 · 4 citations