Assesses optical and mechanical changes in intraocular lenses under UV exposure, suggesting impacts on lens quality.
Hydrophobic acrylic intraocular lenses (IOLs) undergo significant optical, chemical, surface, and mechanical changes upon prolonged ultraviolet (UV) exposure. We assessed a commercial hydrophobic acrylic IOL using an accelerated UV protocol (1–25 h) and a multimodal workflow combining UV–Vis spectrophotometry, ATR–FTIR spectroscopy, contact-angle measurements, and Shore C hardness testing. UV–Vis spectroscopy revealed up to a 40 % decrease in UV transmittance after 25 h, and a 36–50 % increase in absorbance extending into the short-wavelength visible region, consistent with the formation of additional absorbing centers that may influence long-term optical quality. ATR-FTIR analyses indicated photochemical modifications in the acrylate backbone compatible with photo-oxidation, including chain scission and ester rearrangement. Wettability showed a transient hydrophobic shift with peaks at approximately 3 h for both test liquids, with a more pronounced hydrophobic shift for the glucose solution, followed by a return to baseline by 25 h, suggesting competing surface reorganization and oxidation processes. Shore C hardness increased from 63.0 to 73.1 after 25 h, indicating UV-induced stiffening. To our knowledge, this is the first study to simultaneously report these four modalities together for accelerated UV aging of a single hydrophobic acrylic IOL model.
No takes yet. Share an insight, caveat, or question.
Torrisi et al. (2026) studied this question.