Why the study?
The range and limits of human transthoracic impedance, along with potential influencing factors, are of high interest for cardiac defibrillator testing and design.
What is the range of human transthoracic impedance during cardiac defibrillation and what factors influence it?
What is the range of human transthoracic impedance during cardiac defibrillation and what factors influence it?
Human transthoracic impedance during defibrillation averages 76.7 Ω but varies widely, primarily influenced by coupling device, electrode size, and electrode pressure.
Answers are grounded in this paper and cite their sources.
Alerts clinicians to wide TTI variability in defibrillation; extends evidence on modifiable factors like electrode size and pressure.
For cardiac defibrillator testing and design purposes, the range and limits of the human TTI is of high interest. Potential influencing factors regarding the electronic configurations, the electrode/tissue interface and patient characteristics were identified and analyzed. A literature survey based on 71 selected articles was used to review and assess human TTI and the influencing factors found. The human TTI extended from 12 to 212 Ω in the literature selected. Excluding outliers and pediatric measurements, the mean TTI recordings ranged from 51 to 112 Ω with an average TTI of 76.7 Ω under normal distribution. The wide range of human impedance can be attributed to 12 different influencing factors, including shock waveforms and protocols, coupling devices, electrode size and pressure, electrode position, patient age, gender, body dimensions, respiration and lung volume, blood hemoglobin saturation and different pathologies. The coupling device, electrode size and electrode pressure have the greatest influence on TTI.
No takes yet. Share an insight, caveat, or question.
Heyer et al. (2022) studied this question.