Oxide films grown anodically at the surface of superpure aluminium are used as standards to assess the accuracy of the thickness ( d XPS ) determined using the Beer–Lambert treatment of the Al 2p metal and oxide peak intensities. For the fitting conditions employed, the value of d XPS is found to be very close to the true film thickness for films <10 nm thick, beyond which imprecise values are obtained. A surface layer of hydration is identified from the curve fitting of the O 1s core level, which would qualitatively account for the slight underestimate of film thickness provided by the expression for d XPS . The validity of the Al 2p fit may be determined through this correlation of thickness but requires first a quantitative assessment of the thickness and composition of the hydroxide‐rich layer.