Evaluation reveals accuracy of the Lifestat 1600 Pulse Oximeter in measuring oxygen levels, suggesting clinical reliability.
This study evaluated the accuracy of the SaO 2 measured by the Physio-Control Lifestat 1600 Pulse Oximeter. Method: We evaluated the accuracy of this oximeter in the SaO 2 range of 75-100% in 9 normal, healthy, Caucasian volunteers, who had blood-sampling catheters placed in their radial arteries. Hypoxemia was produced by a rebreathing technique, normoxemia by having the subjects breathe room air, and hyperoxemia by having them breathe supplemental oxygen. After the subject's SaO 2 stabilized at the desired level, a pulse oximeter SaO 2 reading was taken and a simultaneously drawn arterial blood sample was analyzed on an IL282 CO-Oximeter for SaO 2 , carboxyhemoglobin (COHb), and methemoglobin (metHb). Results: In the 114 data sets collected, the mean (± SD) Physio-Control SaO 2 was 87.9 ± 6.6% and the mean IL282 SaO 2 was 88.2 ± 6.4% (significantly different: P = 0.03). The mean COHb was 2.5 ± 0.5%, and the mean metHb was 0.5 ± 0.2%. There was a 0.97 correlation between the Physio-Control and the IL282-SaO 2 values. The limits of agreement for the Physio-Control SaO 2 were ± 3%. That is, the Physio-Control SaO 2 tended to be within 3 percentage points of the IL282 SaO 2 . Conclusions: Although the difference between the mean SaO 2 values produced by the pulse oximeter and the CO-Oximeter was statistically significant, we consider it too small to be clinically important. When COHb and metHb levels are low, as they were in this normal population, the Physio-Control Lifestat 1600 Pulse Oximeter is acceptably accurate in the SaO 2 range of 75-100%.
No takes yet. Share an insight, caveat, or question.
Hess et al. (1987) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: