Key result
Pulse oximetry accurately estimated arterial oxygen saturation during exercise in healthy subjects, showing strong correlations and insignificant differences versus CO-oximetry.
Why the study?
Does pulse oximetry accurately estimate arterial oxygen saturation during exercise in healthy subjects compared to direct arterial blood sampling?
Cross-Sectional (n=10)
Does pulse oximetry accurately estimate arterial oxygen saturation during exercise in healthy subjects compared to direct arterial blood sampling?
Pulse oximetry provides a useful and reasonably accurate estimation of arterial oxygen saturation during exercise in healthy subjects.
Answers are grounded in this paper and cite their sources.
May support pulse oximetry in healthy exercise research; leaves open validation in patients.
The accuracy of two pulse oximeters (Ohmeda 3700 and Biox IIa) was evaluated during cycle ergometer incremental exercise in 10 healthy subjects. The exercise protocol began at 30 W with the power output being increased 15 W.min-1 until volitional fatigue. Ear and finger probe pulse oximetry measurements of available hemoglobin (%Spo2) were compared with arterial oxyhemoglobin fraction of total hemoglobin (%HbO2) measured directly from arterial blood samples using a CO-oximeter. To provide a wide range of %HbO2 values, four subjects exercised under hypoxic conditions [inspired partial pressure of O2 (PIo2) = 107 Torr], while the remaining six subjects exercised under normoxic conditions (PIo2 = 150 Torr). Because carboxyhemoglobin (HbCO) or methemoglobin (MetHb) is not measured by pulse oximeters, %HbO2 was corrected for HbCO and MetHb and expressed as percent arterial O2 saturation of available Hb (%Sao2). Small and insignificant differences (P greater than 0.05) existed between SpO2 (all 3 instruments) and %SaO2 at the lowest work rate and the highest power output achieved. Regression analyses of %SpO2 vs. %SaO2 produced correlation coefficients of r = 0.82 [standard error of the estimate [(SEE) = 1.79], r = 0.89 (SEE = 1.48), and r = 0.93 (SEE = 1.14) for the Biox IIa, Ohmeda 3700 (ear), and the Ohmeda 3700 (finger) pulse oximeters, respectively. We conclude that pulse oximetry, within the above limits of accuracy, is useful in estimating %SaO2 during exercise in healthy subjects.
No takes yet. Share an insight, caveat, or question.
Powers et al. (1989) conducted a cross-sectional in Healthy subjects (n=10). Pulse oximetry (Ohmeda 3700 and Biox IIa) vs. Arterial blood samples using a CO-oximeter was evaluated on Correlation between pulse oximetry (%SpO2) and arterial O2 saturation (%SaO2) (r = 0.82 to 0.93, p=>0.05). Pulse oximetry accurately estimated arterial oxygen saturation during exercise in healthy subjects, showing strong correlations and insignificant differences versus CO-oximetry.