Answers are grounded in this paper and cite their sources.
A multiparameter gas sensor based on distributed feedback interband cascade lasers emitting at 4.35 μm and ultrafast electro-spun luminescence oxygen sensors has been developed for the quantification and continuous monitoring of ¹³CO₂/¹²CO₂ isotopic ratio changes and oxygen in exhaled mouse breath samples. Mid-infrared absorption spectra for quantitatively monitoring the enrichment of ¹³CO₂ levels were recorded in a miniaturized dual-channel substrate-integrated hollow waveguide using balanced ratiometric detection, whereas luminescence quenching was used for synchronously detecting exhaled oxygen levels. Allan variance analysis verified a CO₂ measurement precision of 1.6‰ during a 480 s integration time. Routine online monitoring of exhaled mouse breath was performed in 14 mechanically ventilated and instrumented mice and demonstrated the feasibility of online isotope-selective exhaled breath analysis within microliters of probed gas samples using the reported combined sensor platform.
No takes yet. Share an insight, caveat, or question.
Tütüncü et al. (2018) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: