Why the study?
Patient-device relationships between TAVR and BAV differ from those in TAV, but the nature of these differences across common BAV morphologies is not well understood.
Does bicuspid aortic valve morphology affect TAVR device expansion, alignment, and stress compared to tricuspid aortic valves?
Does bicuspid aortic valve morphology affect TAVR device expansion, alignment, and stress compared to tricuspid aortic valves?
Finite element simulation demonstrates that TAVR deployment in bicuspid aortic valves results in significantly higher leaflet and aortic wall stress, as well as greater device asymmetry, compared to tricuspid valves, with variations depending on the BAV subtype.
Answers are grounded in this paper and cite their sources.
Suggests caution for TAVR in BAV; hypothesis-generating for morphology-specific device trials.
The bicuspid aortic valve (BAV) is a common and heterogeneous congenital heart abnormality that is often complicated by aortic stenosis. Although initially developed for tricuspid aortic valves (TAV), transcatheter aortic valve replacement (TAVR) devices are increasingly applied to the treatment of BAV stenosis. It is known that patient-device relationship between TAVR and BAV are not equivalent to those observed in TAV but the nature of these differences are not well understood. We sought to better understand the patient-device relationships between TAVR devices and the two most common morphologies of BAV. We performed finite element simulation of TAVR deployment into three cases of idealized aortic anatomies (TAV, Sievers 0 BAV, Sievers 1 BAV), derived from patient-specific measurements. Valve leaflet von Mises stress at the aortic commissures differed by valve configuration over a ten-fold range (TAV: 0.55 MPa, Sievers 0: 6.64 MPa, and Sievers 1: 4.19 MPa). First principle stress on the aortic wall was greater in Sievers 1 (0.316 MPa) and Sievers 0 BAV (0.137 MPa) compared to TAV (0.056 MPa). TAVR placement in Sievers 1 BAV demonstrated significant device asymmetric alignment, with 1.09 mm of displacement between the center of the device measured at the annulus and at the leaflet free edge. This orifice displacement was marginal in TAV (0.33 mm) and even lower in Sievers 0 BAV (0.23 mm). BAV TAVR, depending on the subtype involved, may encounter disparate combinations of device under expansion and asymmetry compared to TAV deployment. Understanding the impacts of BAV morphology on patient-device relationships can help improve device selection, patient eligibility, and the overall safety of TAVR in BAV.
No takes yet. Share an insight, caveat, or question.
Kusner et al. (2021) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: