Key result
SMART Pass filtering is linked to ~33% fewer inappropriate shocks in S-ICD patients.
Why the study?
Use of the subcutaneous implantable cardioverter defibrillator (S-ICD) is increasing, but evidence from industry-independent and unselected populations remains limited.
Observational (n=4,924)
Yes
Answers are grounded in this paper and cite their sources.
In a large, nationwide, unselected French cohort, S-ICD therapy demonstrated a 5-year inappropriate shock rate of 13.8% and early battery depletion in 10.8%, providing real-world long-term safety data.
BACKGROUND AND AIMS: Although use of the subcutaneous implantable cardioverter defibrillator (S-ICD) is increasing, evidence from industry-independent and unselected populations remains limited. METHODS: HONEST is a ongoing nationwide academic observational study enrolling 98.2% of patients implanted with an S-ICD across France (2012-2019). Five-year clinical endpoints were centrally adjudicated. RESULTS: Overall, 4924 patients were enrolled (mean age 49.9 ± 15 years, 76.7% male, 63.0% for primary prevention). Implants used general anaesthesia (78.9%), and defibrillation testing (82.6%). Perioperative complications (within 30 days) occurred in 4.4%. At 5 years, cumulative incidence rates were 13.8% for inappropriate shocks, 10.8% for early battery depletion, 2.4% for infections, 1.5% for lead dysfunction, and 1.4% for chronic discomfort. Reoperation was required in 16.9%, need for cardiac pacing in 3.1%, and definite S-ICD extraction in 8.4%. Inappropriate shocks were independently associated with male sex (hazard ratio [HR] 1.29, 95% confidence interval [CI] 1.14-1.46, P < .001), obesity (HR 1.35, 95% CI 1.02-1.79, P = .032), arrhythmogenic right ventricular cardiomyopathy (HR 1.70, 95% CI 1.03-2.81, P = .036), and the presence of a pacemaker (HR 2.20, 95% CI 1.16-4.17, P = .016). SMART Pass filtering significantly reduced inappropriate shocks (HR 0.67, 95% CI 0.50-0.89, P = .007). Among patients with inappropriate shocks, ∼1% developed induced ventricular fibrillation (one fatality), and 10% underwent device extraction. Ineffective shocks or undetected arrhythmias occurred in only 0.2%. Among 547 deaths (11.1%), 53.9% were cardiovascular, including 26 sudden deaths, and 8 were S-ICD/procedure-related, with none related to S-ICD extraction. CONCLUSIONS: This nationwide study refines the long-term event profile of S-ICD therapy and may inform clinical practice and device selection.
No takes yet. Share an insight, caveat, or question.
Kerkouri et al. (2025) conducted an observational in Subcutaneous implantable cardioverter defibrillator (S-ICD) implantation (n=4,924). Subcutaneous implantable cardioverter defibrillator (S-ICD) was evaluated on Inappropriate shocks at 5 years. In a nationwide French cohort of 4,924 patients with an S-ICD, the 5-year cumulative incidence of inappropriate shocks was 13.8%, which was reduced by SMART Pass filtering.
Synapse has enriched 10 related papers. Consider them for comparative context: